Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values for four computed tomography scanners

Yada, Nobuhiro and Onishi, Hideo (2016) Validation of computed tomography-based attenuation correction of deviation between theoretical and actual values for four computed tomography scanners. Asia Oceania Journal of Nuclear Medicine and Biology, 4 (2). pp. 81-89.

[img]
Preview
Text
AOJNMB_Volume 4_Issue 2_Pages 81-89.pdf

Download (1MB) | Preview
Official URL: http://aojnmb.mums.ac.ir/article_6784.html

Abstract

Objective: In this study, we aimed to validate the accuracy of computed tomography-based attenuation correction (CTAC) using the bilinear scaling method.Methods: The measured attenuation coefficient (μm) was compared to a theoretical attenuation coefficient (μt ) using four different CT scanners and an RMI 467 phantom. The effective energy of the CT beam X-rays was calculated, using the aluminum half-value layer method, and was used in conjunction with an attenuation map to convert the CT numbers to μm values for the photon energy of 140 keV. We measured the CT number of the RMI 467 phantom for each of four scanners, and compared the μm and μt values for the effective energies of the CT beam X-rays, effective atomic numbers, and physical densities.Results: The μm values for CT beam X-rays with low effective energies decreased in high construction elements, compared with CT beam X-rays of high effective energies . As the physical density increased, the μm values elevated linearly. Compared with other scanners, the μm values obtained from the scanner with CT beam X-rays of the maximal effective energy increased once the effective atomic number exceeded 10.00. The μm value of soft tissue was equivalent to the μt value. However, the ratios of the maximal differences between the μm value and the μt value were 25.4 (lung) and 21.5 (bone) respectively. Additionally, the maximal differences in the μm values were 6.0 in the bone tissue for each scanner.Conclusion: The bilinear scaling method could accurately convert CT numbers to μ values within the soft tissues.

Item Type: Article
Subjects: WN Radiology . Diagnostic Imaging
Divisions: Journals > Asia Oceania J Nuclear Medicine & Biology
Depositing User: AOJNMB AOJNMB
Date Deposited: 19 Sep 2017 12:53
Last Modified: 19 Sep 2017 12:53
URI: http://eprints.mums.ac.ir/id/eprint/1011

Actions (login required)

View Item View Item