Assessment of the effects of radiation type and energy on the calibration of TLD-100

Bahreyni Toossi, Mohammad Taghi and Gholamhosseinian, Hamid and Vejdani Noghreiyan, Atefeh (2018) Assessment of the effects of radiation type and energy on the calibration of TLD-100. Iranian Journal of Medical Physics, 15 (3). pp. 140-145.

[img] Text
Vejdani02.pdf

Download (1MB)
Official URL: http://ijmp.mums.ac.ir/article_9933.html

Abstract

Introduction: In radiation therapy, knowing the dose rates to healthy organs and tumors is beneficial, and thermoluminescent dosimeter (TLD) allows for this possibility. This study was aimed at determining the dose-response differences of TLDs in various types of radiation, energy levels, and dose rate calibrated with other types of radiation beams and energy and dose levels. Materials and Methods: In this study, LiF:Mg,Ti (TLD-100) was used for dosimetry. Photon and electron irradiation was performed by Elekta Precise Linear Accelerator. First, TLDs were calibrated in three different groups of 6 MV photon, 6 MeV electron, and 60Co teletherapy photon beam with 50 cGy dose. Next, each group was irradiated with 6 MV photon, 6 MeV electron, and 60Co teletherapy photon beam separately at three different dose levels of 20, 60, and 100 cGy. Results: TLDs calibrated with electron were significantly different at all dose levels and with all types of radiation from TLDs calibrated with photon or 60Co teletherapy photon beam (P=0.000). P-value of the TLDs calibrated with 6 MV photon versus 60Co was less than 0.94. The maximum standard deviation belonged to 100 cGy irradiation, while the least pertained to 20 cGy irradiation. Conclusion: Calibration of TLDs depends on the type of radiation.

Item Type: Article
Subjects: WN Radiology . Diagnostic Imaging
Divisions: Journals > Iranian J Medical Physics
Depositing User: ijmp ijmp
Date Deposited: 27 Jun 2018 08:35
Last Modified: 27 Jun 2018 08:35
URI: http://eprints.mums.ac.ir/id/eprint/9861

Actions (login required)

View Item View Item